검색 전체 메뉴
PDF
맨 위로
OA 학술지
Post-tuning of Sample Position in Common-path Swept-source Optical Coherence Tomography
  • 비영리 CC BY-NC
  • 비영리 CC BY-NC
ABSTRACT
Post-tuning of Sample Position in Common-path Swept-source Optical Coherence Tomography
KEYWORD
Optical coherence tomography , Common-path interferometer , Biological imaging
  • I. INTRODUCTION

    Optical coherence tomography (OCT) is an emerging technology for non-invasive cross-sectional imaging of biomedical tissue samples [1]. It has been investigated for use in numerous applications, and thus, many types of system configurations have been developed. In most OCT systems, two beams are propagated to the reference arm and the sample arm and then re-coupled to produce an interfered optical signal [2,3]. The general configuration of these two separate arms has limited the application of OCT systems in endoscopic imaging because an arbitrary probe length is required to deliver the optical signal from the internal organ tissue to the optical imaging system, which exists outside the body.

    Recently, a common-path OCT system based on fiber-optic autocorrelation has been developed in order to overcome the length, polarization, and dispersion mismatch between the reference and sample arms [4-7]. Since there is only one path in common-path OCT, the reference and sample signals are reflected separately from the partial reflecting probe and the sample, respectively, and they propagate back through a commonly shared optical path [8]. However, a critical problem with conventional common-path OCT is that the sample position distance between the reference and sample positions is limited. The interfered optical signal between the reference and sample signals can be detected only when the delay time is within the coherence time of the light source. If the delay time between the reference and sample signals is τS and the coherence time of the light source is τC, τS should be shorter than τC to produce an interfered optical signal. This means that the sample should be located inconveniently close to the reference position. As a result, unlike in the case of conventional Michelson-and Mach-Zehnder-type OCT, an objective lens with a long working distance cannot be utilized, and in-depth information about the sample cannot be obtained in this common-path OCT configuration. In the general OCT system, an objective lens with a low numerical aperture and a long working distance is commonly used to increase the penetration depth.

    In this research, we developed a novel method for post-tuning the optical path-length difference (OPD) between the reference and sample signals for an arbitrary location of the reference and sample positions [4]. We present a novel common-path OCT configuration by including an additional Mach-Zehnder interferometer in the post-path of the interfered optical signal. Both 1-D profiles and 2-D images obtained with this novel common-path OCT and the conventional common-path OCT are compared. The OCT image is implemented with a Fourier domain OCT (FD-OCT) system based on a wavelength-swept laser source and a balanced detector [9-13].

    II. TUNING OF THE OPTICAL PATH LENGTH

    Fig. 1 shows a schematic of the novel common-path OCT system, in which the OPD is tuned by adding an extra interferometer in the post stage. A light beam from the wavelength-swept laser source propagates from the first port to the second port of the circulator and is transmitted through a probe fiber. The focused beam is reflected from the single-faced partial reflector and the sample, respectively. Then, it propagates to the third port of the circulator. Only one face of the partial reflector induces a small portion of the reflected light to act as the reference signal for the interferometer. The reference and sample signals share the common path of the fiber probe. There is a time delay, τS, between the two signals from the single-face partial reflector and the sample. Assuming that the reference signal and sample signal are ER and ES, respectively, the electric field E3-port(t) at the third port of the circulator is given by

    image

    where RR and RS are the reflectivities of the partial reflector and the sample, respectively.

    In general, the axial imaging depth is determined by the OPD between the reference and the sample. Because of the finite spectral resolution of the spectrometer in spectral-domain (SD) OCT or the instantaneous linewidth of the wavelength-swept laser in swept-source (SS) OCT, an increase

    in the OPD leads to decay of the signal-to-noise ratio (SNR) and limits the range of the axial depth. Thus, in conven-tional common-path OCT, the more closely positioned partial reflector can be interfered to the sample with higher visibility signal. On the other hand, the OCT image also depends on the reflected optical intensity from the sample, which becomes maximal when the sample is located at the focal position of the objective lens. However, this results in a decreased SNR from the sample because the distance at the focal position of the objective lens corresponds to the increased OPD in the common-path OCT system.

    To solve the above dilemma of the sample position, whether it be closer to the partial reflector or closer to the focal position, we suggest the addition of a Mach-Zehnder interferometer including a variable coupler, a tunable optical delay line (ODL), and a 50:50 coupler in the post-path of the interfered optical signal, as shown in Fig. 1. The signal from the third port of the circulator enters a variable coupler and propagates separately in two fiber paths (①, ②). Then, those are combined at a 50:50 coupler. Therefore, the two terms on the right-hand side of equation (1) become four terms. When the splitting ratio of the variable coupler is 50%, the electric field along paths ① and ② can be mathematically expressed as

    image

    where τD denotes the additional delay time introduced by a tunable ODL.

    Therefore, we can show that the total electric field detected at one of the balanced photodiodes becomes

    image

    Among the combined terms with the reference and sample signals in the curly braces, it is clear that the term

    image

    can be the final interference signal in the case of

    image

    This means that the post-tuning of the OPD in the additional Mach-Zehnder interferometer can induce the minimum OPD condition even when the sample is located at the focal position of the objective lens.

    To determine the effect of the additional Mach-Zehnder interferometer, we can analyze the SNR of our proposed system. From the previous reports, SNR is defined as the signal power divided by the noise process variance [5,14-16].

    image

    where

    image

    means the mean square signal photocurrent and σ2 is the total noise variance.

    Using each mean photocurrent and noise variance in a conventional system and in our system, our proposed system shows theoretically 6 dB lower sensitivity than the conventional system [5,14-16]. However, our system is still promising because it provides flexible use of common-path probes by post-tuning of sample position distance. It means that this method is more suitable to obtain a clear cross sectional sample image at the long distance of sample position.

    III. EXPERIMENTAL CONDITION

    In this swept source OCT system, the output spectra from the wavelength-swept laser used is shown in Fig. 2(a). The laser has a fiber ring cavity configuration, with a semiconductor amplifier (SOA, InPhenix, Inc.), a piezo-actuated fiber Fabry-Perot tunable filter (FFP-TF, Lambda Quest, Inc.), an isolator, a polarization controller, and a fiber output coupler. The laser

    output has a center wavelength of 1305 nm, a bandwidth of 80 nm, and an output power of 10 mW. As we used a Fourier-domain mode locked (FDML) swept laser with a 2.8 km delayed fiber in the ring cavity, a high repetition rate of 71.09 kHz is operated with stability. Fig. 2(b) shows the measurement result for an axial resolution of the OCT system processed by the Fourier transform of the acquired interferogram. The measured axial resolution is 12.8 μm at the OPD of 1.8 mm, which differs slightly from the theoretical value of 10 μm. It is thought that this difference can be attributed to the effect of dispersion introduced by the additional interferometer.

    The objective lens, (model LSM03, magnification factor = x5, Thorlabs), has a focal length of 35.9 mm and a lens working distance of 25.1 mm. Its transverse resolution is estimated to be 25 μm from the mean spot size at the focal position. The partial reflector consists of two-sided glass. The antireflection-coated side has a reflectivity below 0.002, and the other side has a reflectivity of 0.05 and a transmittance of 0.95 in the 1300 ± 50 nm region. By posttuning the OPD, we could obtain OCT images, regardless of the distance between the partial reflector and the sample, within the long working distance of the objective lens. For the increased range of the sample position distance, the objective lens with lower magnification factor can be helpful to increase the working distance though the transverse resolution will be decreased.

    An interferogram signal is detected by the balanced photodiode (model 1817, New Focus, USA), which has a band-width of 80 MHz, a trans-impedance gain of 50,000 V/A, and a responsivity of 0.8 A/W at 1300 nm. By dual balancing, the DC level is suppressed and no additional background subtraction is conducted. Acquired interferogram data are trans-ferred to a personal computer via a data-acquisition board (PCI-5124, National Instruments Corp., USA), which has a maximum sampling rate of 200 Msamples/s for the analog signal. Fast Fourier transformation is performed on the transferred data. To acquire a single cross-sectional image, a linear translation stage is driven with a saw-tooth waveform to provide transverse scanning over a 3 mm distance along the sample surface. General LabVIEW software is used to process the image.

    IV. EXPERIMENTAL COMPARISON

    To discriminate between the absolute distance (sample position distance) and the relative distance (OPD) under the same conditions, we fixed the location of the partial reflector at 6 mm from the focal position, as shown in Fig. 1, by the repeated optimization process of the reference position for the maximized SNR [4]. Within the lens working distance of the objective lens, which is 25.1 mm, the position of the mirror sample is changed along the axial beam distance to measure a series of point-spread functions (PSFs). We call this distance between the focal position and the sample surface the “off-focal distance.”

    First, for the conventional common-path OCT, the dynamic range roll-off is measured in Fig. 3 as the off-focal distance is changed incrementally. The off-focal distance range 0-9 mm corresponds to both “sample position distance” and “OPD” range of 6-15 mm because the interfered optical signal by the OPD is exactly the same with the sample position distance between the partial reflector and the sample in this conventional common-path OCT. The dynamic range is determined by the ratio of the PSF peak value and the noise level. The noise level is determined to be -5 dB by averaging the region between 0 and 9 mm. The system dynamic range drops rapidly Fig. 3(a), the dynamic range drops rapidly from 35.9 to 7.6 dB by increasing the off-focal distance from 0 to 10 mm. The 6 dB drop out is measured at off-focal distance of around 6 mm. As mentioned in the previous section, this rapid roll-off of 27.9 dB is due to two reasons. First, the finite sample number due to the thick instantaneous linewidth in the swept source leads to strong decay of the SNR for a deeper axial range of the incremental OPD. Second, the signal beam has weak reflectivity as the sample moves away from the focal position [9,10]. Therefore, when imaging with the conventional common-path OCT, the sample should be located as close as possible to the partial reflector at

    around the focal position of the objective lens. These two requirements may restrict the available position of the sample significantly in the limited range.

    In Fig. 3(b), for the purpose of comparison, the dynamic range is measured again using the novel common-path OCT setup by locating the mirror sample at the same off-focal distances from 0 to 9 mm. At every off-focal distance, the extra delay time, τD, of the fiber path ② can be manually tuned to be the same as the original delay time, τS, and thus, the PSF peak can be set at the OPD near 0 mm repeatedly. The recovery of the dynamic range is clearly monitored from the control of the OPD so that the dynamic range drops slightly from 36.9 to 34.2 dB when the off-focal distance is increased from 0 to 10 mm. It seems that this slight roll-off of 2.7 dB is due to one reason: the sample signal has weak reflectivity when the sample is away from the focal position. It also means that the effect of the focal position on the quality of the OCT image is relatively weaker than that of the OPD. Moreover, the effect of the OPD can be ignored because the sample position distance between the partial reflector and the sample is compensated by tuning the path length by τD. This shows that our novel common-path OCT system can be just as useful as the other two separate-path OCT systems on the distance flexibility of sample position. We expect the slight fluctuation of PSF peak values in Fig. 3 can be suppressed when we further improve the axial alignment of optical components and the feedback control of the manual OPD compensation.

    Dynamic range can also be compared for various samples by using the conventional and novel common-path OCT systems. Fig. 4 shows the depth profiles of a sample consisting of two slide glasses acquired at different sample position distances between the partial reflector and the sample using a conventional common-path OCT system. When the top surface of the sample is positioned close to the partial reflector at 0.3 mm (Fig. 4(a)), we can observe the two glasses and the air gap between them clearly. As the sample position distance increases, the image quality is strongly degraded because of the dynamic range roll-off. When the distance between the partial reflector and the top surface of the sample becomes more than 3.3 mm, the top surface of the upper glass can be monitored with difficulty, the other surfaces not at all. It means that there is no way to get a good-quality image in the conventional common-path OCT system without positioning the sample close to the partial reflector.

    For the additional comparison with a cross sectional OCT imaging, the biological sample of a fish eye (Pangasius sutchi) is also imaged in our proposed swept-source common-path OCT system by changing the sample position distance, as shown in Fig. 5. For the much extended range of sample position distance, we used the other objective lens, (model LSM04, magnification factor = x3, Thorlabs), with a focal length of 53.9 mm and a lens working distance of 42.3 mm. When the sample surface was placed at the focal position and the partial reflector was located at 6 mm from the

    focal position as shown in Fig. 5 (a), the inner structure of eye sample can be clearly observed as shown in Fig. 5 (b). For the estimate the maximum available range of sample position distance, we moved the partial reflector toward the objective lens and the sample in the opposite direction, simultaneously, up to 10 mm for both sides. As the sample moves away from the reference position, we also increase the extra delay time, τD, to move the highest PSF peak toward the 0 mm point of the OPD. As shown in Fig. 5 (c) at the sample position distance of 26 mm, the quality of OCT image becomes degraded because the beam intensity and transverse resolution are relatively decreased due to the long distance from the focal position. The effect of intensity degradation is as low as 2.7 dB for the off-focal distance of 10 mm, as shown in Fig. 3 (b). The mean spot size of this objective lens is 35 μm at the focal position and the transverse resolution can be estimated to ~ 100 μm at the off-focal distance of 10 mm. This means that there is a negligible degradation in the axial image quality up to the sample position distance of 26 mm, as predicted in Fig. 3(b), and the distance problem between the partial reflector and the sample can be solved for the practical endoscopic applications. The image size is 3 mm × 3 mm, which consists of 256 axial scans (transverse pixels) with 256 points (axial pixels) per scan. In case of a conventional common-path OCT system, as expected from the results of strong fall-off in Fig. 4, it is monitored that image quality is seriously degraded at a few mm of the sample position distance. Moreover optical interference does not occur at more than ~ 20 mm of sample position distance.

    V. CONCLUSION

    We have developed a novel method for post-tuning the sample position distance in a common-path OCT system by including an additional Mach-Zehnder interferometer in the post-path of an interfered optical signal. From the experimental results of post-imaging depth tuning, the dynamic roll-off of the system is reduced to be as low as 2.7 dB during 10 mm axial scanning. This can solve the positioning problem for the objective lens, partial reflector, and sample in the conventional common-path OCT system. The proposed method may allow for greater flexibility in the use of OCT systems in endoscopic applications.

참고문헌
  • 1. Huang D., Swanson E., Lin C., Schuman J., Stinson W., Chang W., Hee M., Flotte T., Gregory K., Puliafito C., Fujimoto J. (1991) “Optical coherence tomography” [Science] Vol.254 P.1178-1181 google cross ref
  • 2. Youngquist R. C., Carr S., Davies D. E. N. (1987) “Optical coherence-reflectometry: a new optical evaluation technique” [Opt. Lett.] Vol.12 P.158-160 google cross ref
  • 3. Takad K., Yokohama I., Chida K., Noda J. (1987) “New measurement system for fault location in optical waveguide devices based on an interferometric technique” [Appl. Opt.] Vol.26 P.1603-1606 google cross ref
  • 4. Sharma U., Fried N. M., Kang J. U. (2005) “All-fiber common-path optical coherence tomography: sensitivity optimization and system analysis” [IEEE J. Select. Topics Quantum] Vol.11 P.799-805 google cross ref
  • 5. Kang J. U., Rodrigues A. 2007 “Fourier domain common-path fiber OCT with tunable reference: analysis and optimization” [in Proc. CLEO-QELS 2007] google
  • 6. Vakhtin A. B., Kane D. J., Wood W. R., Peterson K. A. (2003) “Common-path interferometer for frequency-domain optical coherence tomography” [Appl. Opt.] Vol.42 P.6953-6958 google cross ref
  • 7. Tumlinson A. R., Barton J. K, Povazay B., Sattman H., Unterhuber A., Leitgeb R. A., Drexler W. (2006) “Endoscope-tip interferometer for ultrahigh resolution frequency domain optical coherence tomography in mouse colon” [Opt. Express] Vol.14 P.1878-1887 google cross ref
  • 8. Park J. S., Jeong M. Y., Jung C. H., Ouh C. H., Kang H. J., Han Y. G., Lee S. B., Kim C. S. (2010) “Flexible curled optical cord for bending-insensitive optical imaging delivery” [IEEE J. Select. Topics Quantum Electron.] Vol.16 P.1031-1038 google
  • 9. Yun S. H., Tearney G. J., de Boer J. F., Bouma B. E. (2004) “Removing the depth-degeneracy in optical frequency domain imaging with frequency shifting” [Opt. Express] Vol.12 P.4822-4828 google cross ref
  • 10. de Boer J. F., Cense B., Park B. H., Pierce M. C., Tearney G. J., Bouma B. E. (2003) “Improved signal to noise ratio in spectral-domain compared with time-domain optical coherence tomography” [Opt. Lett.] Vol.28 P.2067-2069 google cross ref
  • 11. Nakamura Y., Makita S., Yamanari M., Itoh M., Yatagai T., Yasuno Y. (2007) “High-speed three-dimensional human retinal imaging by line-field spectral domain optical coherence tomography” [Opt. Express] Vol.15 P.7103-7116 google cross ref
  • 12. Lee J. H., Jung E. J., Kim C. S. (2010) “Optical coherence tomography based on a continuous-wave supercontinuum seeded by erbium-doped fiber’s amplified spontaneous emission” [J. Opt. Soc. Korea] Vol.14 -49 google cross ref
  • 13. Huber R., Wojtkowski M., Taira K., Fujimoto J. G. (2005) “Amplified, frequency swept lasers for frequency domain reflectometry and OCT imaging: design and scaling principles” [Opt. Express] Vol.13 P.3513-3528 google cross ref
  • 14. Leitgeb R., Hitzenberger C. K., Fercher A. F. (2003) “Perfor-mance of fourier domain vs. time domain optical coherence tomography” [Opt. Express] Vol.11 P.889-894 google cross ref
  • 15. Rollins A. M., Izatt J. A. (1999) “Optimal interferometer designs for optical coherence tomography” [Opt. Lett.] Vol.24 P.1484-1486 google cross ref
  • 16. Choma M. A., Sarunic M. V., Yang C., Izatt J. A. (2003) “Sensitivity advantage of swept source and Fourier domain optical coherence tomography” [Opt. Express] Vol.11 P.2183-2189 google cross ref
이미지 / 테이블
  • [ FIG. 1. ]  Schematic of the novel common-path OCT system for tuning the optical path length.
    Schematic of the novel common-path OCT system for tuning the optical path length.
  • [ FIG. 2. ]  (a) Output spectra from the FDML wavelength-swept laser; (b) axial resolution of the OCT system.
    (a) Output spectra from the FDML wavelength-swept laser; (b) axial resolution of the OCT system.
  • [ FIG. 3. ]  Dynamic range determined by locating the mirror sample at various off-focal distances (a) without post-tuning the OPD and (b) by setting the PSF peak at the OPD near 0 mm and tuning τD to be the same as τS.
    Dynamic range determined by locating the mirror sample at various off-focal distances (a) without post-tuning the OPD and (b) by setting the PSF peak at the OPD near 0 mm and tuning τD to be the same as τS.
  • [ FIG. 4. ]  Depth profiles of two slide glasses obtained with the conventional common-path OCT system for different sample position distances: (a) 0.3 mm, (b) 2.3 mm, and (c) 3.3 mm.
    Depth profiles of two slide glasses obtained with the conventional common-path OCT system for different sample position distances: (a) 0.3 mm, (b) 2.3 mm, and (c) 3.3 mm.
  • [ FIG. 5. ]  (a) Setup for cross-sectional images of a fish eye obtained with the proposed common-path OCT system for different sample position distances and the OCT images for the sample position distance of (b) 6 mm and (c) 26 mm, respectively.
    (a) Setup for cross-sectional images of a fish eye obtained with the proposed common-path OCT system for different sample position distances and the OCT images for the sample position distance of (b) 6 mm and (c) 26 mm, respectively.
(우)06579 서울시 서초구 반포대로 201(반포동)
Tel. 02-537-6389 | Fax. 02-590-0571 | 문의 : oak2014@korea.kr
Copyright(c) National Library of Korea. All rights reserved.