This study was performed to evaluate objective visual quality and ocular scattering in myopic refractive-surgery patients, compared to emmetropes. Optical vision-quality parameters (modulation transfer function (MTF) cutoff and Strehl ratio) and objective scattering index (OSI) were measured using an optical quality analysis system (OQAS II) based on the double-pass technique. In all subjects, the higher the MTF cutoff and Strehl ratio, the lower the OSI and ocular higher-order aberrations (HOAs). The MTF cutoff and Strehl ratio for the laser-assisted subepithelial keratectomy (LASEK) group were lower than those for the emmetropia group, while the OSI, ocular HOAs, and spherical aberration (SA) for the LASEK group were higher than those for emmetropia group. Ocular scattering would be one of the important factors in regard to visual quality. Therefore, the quality of the retinal image in the LASEK patients has been shown to reduce the quality of vision more than in the emmetropes.
Laser refractive surgery, as a method to correct refractive errors, has been extensively practiced because of its advantage to maintain good vision with the naked eye, compared to optical correction using glasses or contact lenses. In Korea, which has a high incidence of myopia, refractive-surgical procedures for myopic correction have developed rapidly over the past 20 years, and laser
After corneal refractive surgery, patients are highly satisfied with daytime life. On the other hand, in a modern lifestyle in which city night life increases, it is also true that quality of vision under low lighting conditions is deteriorated, due to decrease of contrast sensitivity, glare, and light blurring (Fig. 1) [1]. These phenomena in the human eye are related to intraocular scattering and optical aberrations, resulting in the degradation of optical quality in visual function [2, 3].
The advent of the Hartmann-Shack wavefront sensor and the double-pass system as a method to evaluate the quality of the retinal image has enabled many researchers to study the effects of higher-order aberrations (HOAs) and intraocular scattering on visual performance, and has provided an opportunity to better understand our visual system. It has been known that the double-pass technique (Fig. 2) [4], which is based on the optical pathway that carries the image on the retina after passage through the ocular media and retinal reflection [5, 6], can provide more accurate estimates or information about retinal image quality [6], as well as being useful for the objective assessment of ocular optical quality [7-9]. The Optical Quality Analysis System (OQAS) is the only instrument with the double-pass technique [6, 10] and is widely used to estimate the quality of the retinal image in clinical practices.
Previous studies have only analyzed ocular scattering to assess optical quality after corneal refractive surgery, or to evaluate visual quality between refractive-surgical techniques, such as LASIK, LASEK, or photorefractive keratectomy (PRK). Since the goal of corneal refractive surgery is to attain emmetropia, comparing the degree of ocular scattering in refractive-surgery patients to that in emmetropes may be helpful in understanding the postoperative ocular optical system, but few studies have been conducted. Therefore, the aim of this study was to compare objective visual quality and ocular scattering based on double-pass retinal images, for myopic refractive-surgery patients who have undergone LASEK and emmetropes.
Subjects were recruited from patients who underwent conventional myopic LASEK, and from emmetropes as a control group. The subjects who met the criterion of best unaided monocular visual acuity of 0.9 (including no more than 0.50D of astigmatic refractive errors) or better in both groups were selected for this study. None of the subjects had any systemic or ophthalmic diseases. This study was approved by the Institutional Review Board (IRB) of Eulji University. All subjects provided written informed consent, following a detailed explanation of the study’s procedures.
2.2. Measurement of Biometric Data
Measurement of uncorrected distance visual acuity (UCDVA) was made under photopic conditions (340 Lx), using high-contrast (100%) ETDRS acuity charts at a distance of 4 m. VA was recorded in logMAR units. In addition, measurement of refractive error was performed by autorefractor (KR-8100P, Topcon, Japan) and phoropter (VT-SE, Topcon, Japan). The testing was carried out monocularly. Corneal thickness was measured by pachymetry function, which is included in the Pentacam Oculyzer topography-measuring system (Oculus Inc., Germany). The natural pupil diameter was taken under scotopic conditions using a digital variable pupillometer (VIPTM- 200, Neuroptics, USA). The pupil diameter can be measured under varying light levels (scotopic, low mesopic, and high mesopic) in one sequence. Measurement was performed three times alternately between right and left eyes.
2.3. Measurement of Optical Parameters
Ocular higher-order aberrations (HOAs) for a 4-mm pupil were analyzed by a wavefront analyzer (KR-1W, Topcon, Japan) with Hartmann-Shack aberrometry, and calculated as the root mean square (RMS) values of the third- and fourth-order Zernike coefficients.
For the quality of the retinal image, optical vision-quality parameters were taken using an optical quality analysis system (OQAS II, Visiometrics, Terrassa, Spain) based on the double-pass technique. The system acquires an image from a point-source object reflecting on the retina, and then directly calculates the modulation transfer function (MTF) from the received double-pass retinal image through Fourier transformation [5]. The MTF represents the contrast loss produced by the ocular optics as a function of spatial frequency, which provides information on the overall ocular optical performance [11].
The OQAS II provides several parameters related to optical quality, including MTF cutoff, Strehl ratio, OQAS values (OVs), and objective scattering index (OSI). The MTF cutoff is the spatial frequency that reaches a value of 0.01 for MTF [9]; the larger the MTF cutoff value, the better the ocular optical quality. The Strehl ratio represents the ratio of peak focal intensity in an aberrated image corresponding to the ideal point-spread function (PSF). The Strehl ratio has a value between 0 and 1.0, with 1.0 indicating an unaberrated, perfect optical system. Therefore, the higher the value of the Strehl ratio, the better the ocular optical quality. Also, the greater the effect of ocular aberrations and scattering, the lower the measured quality of the retinal image [7, 12]. The OVs with 100%, 20%, and 9% contrast levels are standardized values for each spatial frequency corresponding to 0.01, 0.05, and 0.1 MTF values [11, 13]. The higher the OV values, the better the ocular optical quality, and an OV value above 1.0 is associated with higher optical quality [14]. In the case of OVs measured with three contrast levels (100%, 20%, and 9%), OV100% normally has the same value as VA100% (visual acuity), and the values of VA20% and VA9% simulated for OV20% and OV9% were indicated as decimal visual acuity, to express the objective contrast visual quality corresponding to each OQAS value (OV) in the present study. The OSI as an objective parameter is a numerical value obtained by quantifying the degree of intraocular scattered light [9]; the higher the OSI value, the greater the ocular scattering [15, 16].
During the measurements with OQAS II, any spherical refractive error was automatically corrected by the instrument, and cylindrical errors ≥ 0.50 D were corrected with an external trial lens. All measurements were conducted for a 4-mm artificial pupil in mesopic lighting conditions (1 Lx), and were carried out monocularly with an undilated pupil. In addition to this, to minimize the influence of corneal drying, sufficient blinking of the eye was conducted to allow tears to spread sufficiently on the cornea before measurement with OQAS II.
This study involved three visits, with the following procedures conducted at each visit. In the first visit, all subjects underwent an initial optometric examination and were checked according to the inclusion criterion. Also, they received full explanations of the procedures of the study. In the second visit, preoperative biometric data of the subjects were taken. Measurement for biometric data was performed, and followed by that for ocular aberrations. After that, each subject visited another eye hospital for the final measurements using the Pentacam and the OQAS II equipment.
For optical quality parameters, comparisons between the two groups were analyzed by the independent
Subjects were the 88 eyes of 44 patients who underwent conventional myopic LASEK, but not wavefront-guided refractive surgery, and the 40 eyes of 20 emmetropes. The mean uncorrected distance visual acuity (UCDVA) of the subjects was −0.02 ± 0.06 and −0.03 ± 0.05 for LASEK patients and emmetropes respectively. In terms of UCDVA and refractive error (spherical equivalent), there was no statistically significant difference between the two groups. The mean period after refractive surgery was 24.17 ± 17.31 months. The demographics and biometric data of the subjects in the LASEK and emmetropia groups are outlined in Table 1.
[TABLE 1.] Demographics and biometric data of the study’s subjects
Demographics and biometric data of the study’s subjects
The results for optical quality parameters and OSI in the LASEK and emmetropia groups are compared in Table 2. The means for MTF cutoff and Strehl ratio were significantly lower for LASEK (38.377 ± 10.022 and 0.212 ± 0.059 respectively) than those for emmetropes (43.094 ± 8.399 and 0.256 ± 0.058); there were statistically significant differences in MTF cutoff and Strehl ratio between the two groups (
The mean values of optical quality and ocular scattering parameters for a 4-mm artificial pupil, for the LASEK and emmetropia groups
Previous studies performed to compare changes in visual performance outcomes before and after refractive surgery have shown that visual quality was influenced by the increase of ocular HOAs and ocular scattering in corneal-refractive-surgery patients [17-19]. After laser corneal refractive surgery, the main causes of degradation of optical quality are related to increased ocular aberrations [20, 21] and scattering [22-24]. In fact, it is well-known that altered corneal shape and irregular corneal surface due to laser ablation cause these optical defects [25-27].
In the present study, it has been clearly shown that values of OSI and ocular HOAs for the LASEK group were greater than those for the emmetropia group. The results concerning increased ocular HOAs and ocular scattering after refractive surgery were similar to those of previous studies. Additionally, in the emmetropia group the mean values of MTF cutoff, Strehl ratio, and OSI for this study somewhat correspond to those of a previous study, which assessed them for use as reference values for clinical diagnosis in healthy young adults (18 to 30 years) [15]. Thus, the findings imply that the quality of the retinal image for the LASEK group is lower than that for the emmetropes. Meanwhile, the importance of correlations between OSI and achieved refractive correction in LASEK patients have been reported; the results were related to increased ocular scattering by the ablation procedure of refractive surgery [19]. Moreover, Miao
This study also analyzed the correlations between the optical quality parameters (MTF cutoff, Strehl ratio, ocular HOAs, and SA) and objective scatter index (OSI). In the LASEK group, the correlations of MTF cutoff frequency and Strehl ratio with OSI were compared to those of emmetropes (Fig. 3).
The correlations between OSI and MTF cutoff showed a significant negative relationship in both groups: the higher the OSI, the lower the MTF cutoff. The OSI was statistically correlated with MTF cutoff in both groups (
In the LASEK group there was statistical correlation between OSI and ocular HOAs (
As mentioned above, the OVs measured with three contrast levels (100%, 20%, and 9%) were expressed as decimal visual acuity (VA) to objectively compare visual quality between the two groups. The mean values of objective contrast VA with three contrast levels are shown in Table 3. For the LASEK group, the means of VA100%, VA20%, and VA9% were 1.279 ± 0.334, 0.929 ± 0.282, and 0.557 ± 0.176 respectively. For the emmetropia group, the means of VA100%, VA20%, and VA9% were 1.434 ± 0.273, 1.095 ± 0.275, and 0.674 ± 0.169 respectively. As expected, as the contrast decreased, so did the objective visual acuity, in both groups. It was also found that there were statistically significant differences in VA100%, VA20%, and VA9% between the two groups (
The mean values of objective contrast visual acuity corresponding to each OQAS value (Ovs), with contrast levels of 100%, 20%, and 9%, for a 4-mm artificial pupil, in the LASEK and emmetropia groups
The visual acuities of simulated letters “E”, corresponding to retinal images according to values of MTF cutoff and Strehl ratio for the subjects for this study, are shown in Fig. 5. The MTF cutoff values are computed from double-pass retinal images acquired through the ocular media and retinal reflection after point-source object reflection on the retina.
The negative effect of ocular scattering on visual quality has long been studied [24, 28, 29]. Ocular scattering, which significantly deteriorates the quality of the retinal image, is closely related to the point-spread function (PSF; Strehl ratio) as one of the parameters for evaluating optical quality [30]. The PSF represents the distribution of light on the retinal image corresponding to a point source [1]; the greater the distribution of light scattering, the lower the optical quality of the eye. In the present study, the results present examples of different contrast visual acuities according to double-pass images corresponding to the value of MTF cutoff and Strehl ratio in the study groups. As can be seen, the higher the value of MTF cutoff and Strehl ratio, the sharper the image, with less effect of scattering on the retinal image. VAs for three contrast levels and double-pass retinal images corresponding to the means of MTF cutoff and Strehl ratio, in the LASEK group compared to the emmetropia group, are presented in Fig. 6.
It is now accepted as fact that reduction of visual quality after corneal refractive surgery has not been observed when measured using high-contrast visual acuity under photopic conditions. In recent years, low-contrast visual acuity or contrast sensitivity has been commonly used to evaluate the quality of vision after corneal refractive surgery or cataract surgery [31, 32]. Furthermore, it has become possible to explain the loss of contrast caused by deficiency of the ocular optics or contrast sensitivity through a MTF value as a function of spatial frequency [33].
In a normal eye, an MTF cutoff frequency of 30 cpd is considered to be 1.0 in Snellen visual acuity [34]. In this study, it was found that the mean MTF cutoff value was higher than 30 cpd for both groups. Although VA with contrast 100, which could be deemed the maximum visual quality during the day, was greater than 1.0 in both groups, VA100% for the emmetropia group was better than that for the LASEK group. In addition, VA20% and VA9%, which could monitor the degree of loss of visual quality during the night, were also greater than those of the LASEK group. The difference between the two groups was higher in VA9%. These results suggest that the reduction of optical quality in the LASEK group would be more affected than in the emmetropia group, especially under low-contrast conditions.
Ocular scattering should be considered an important factor when evaluating visual quality in refractive-surgery patients. Increased ocular scattering is related to glare, which may cause problems such as difficulty in night driving and photophobia [29, 35]. In particular, it may be difficult to accurately recognize objects under low-illumination conditions [1]. Vilaseca
Ocular scattering is one of the important factors related to visual quality. In all subjects, the higher the MTF cutoff and Strehl ratio, the lower the OSI, ocular HOAs, and SA. For the LASEK group, the MTF cutoff and Strehl ratio were lower than those for the emmetropia group, while the OSI, ocular HOAs, and SA were higher than those for emmetropes. Thus the quality of the retinal image in the LASEK group has been shown to reduce the quality of vision more than in the emmetropia group. In addition, most researchers have compared changes in ocular aberrations and ocular scattering before and after refractive surgery. Thus, comparing visual quality between refractive surgery patients and emmetropes would be helpful for understanding visual performance after refractive surgery, especially at night.