In an optical interferometer that uses sinusoidal modulation and quadrature detection, the amplitude and offset of the interference signal vary with time, even without considering system noise. As a result, the circular Lissajous figure becomes elliptical, with wide lines. We propose and experimentally demonstrate a method for compensating quadrature detection error, based on digital signal processing to deal with scaling and fitting. In scaling, fluctuations in the amplitudes of in-phase and quadrature signals are compensated, and the scaled signals are fitted to a Lissajous unit circle. To do so, we scale the average fluctuation, remove the offset, and fit the ellipse to a unit circle. Our measurements of a target moving with uniform velocity show that we reduce quadrature detection error from 5 to 2 nanometers.
Optical interferometers with quadrature detection have for decades been trusted for displacement measurements that require subnanometer resolution [1-6]. Especially, fiber-based optical interferometers using a telecom wavelength near 1550 nm and a sinusoidally modulated laser for quadrature detection are insensitive to mechanical stress, and benefit from commercialized, small optical components compared to conventional interferometers, which mostly use a 632-nm He-Ne laser and free-space bulk optics [3-6].
The primary performance limitation of fiber-based optical interferometers is quadrature detection error [3-13], which is caused by amplitude differences, dc offsets, and phase delays between the signals [12]. Furthermore, an in sinusoidally modulated optical interferometer, the signal amplitude varies with target movement [2]. Several methods have been introduced to reduce the quadrature detection error. While they showed good results under the condition that the error is fixed during measurement [8-13], in fact phase difference, offset, and gain could change with a moving target. Thus, a real-time compensation method is required.
In this paper, we propose and experimentally demonstrate a simple quadrature-detection-error-compensation method using digital signal processing. The quadrature detection error is mostly due to phase noise, and has the property of being recovered every cycle (period) of the phase signal. Therefore, we can largely reduce such error by compensating the amount of error every cycle for the target’s motion.
A conventional sinusoidally modulated laser interferometer is shown in Fig. 1. The output signal generated by the interfering waves at the photodetector is expressed as [2]
where
where
Eq. (1) can be expanded into
where
From Eqs. (4) and (5) we derive the phase shift
If we move the target from position
However, for quadrature detection error, Eq. (4) and (5) become
where
To reduce this phase error, we propose a calibration method as follows: (1) scale the amplitude fluctuation in the phase signal so that the average amplitude fluctuation is zero, the maximum is +1, and the minimum is −1, and (2) fit the error ellipse to a unit circle using trigonometry. During this process we ignore system noise, including optical nonlinearity due to the high intensity of the light passing through the optical devices, and focus on the quadrature detection error.
After being scaled, the signal
Applying the appropriate trigonometric formulas to and , we obtain Eqs. (12) and (13), which show a 90° phase difference between the two signals, and which simplify to
where
Figure 3 shows the new process, after including our proposed modification. In the scaling section, “[−1: +1]” and “zero offset” denote the normalization process whereby we scale the amplitude to −1 and +1 with an average baseline of zero. In the proposed quadrature section, we fit the input signals sin(
We perform the signal processing using an analog-to-digital converter (ADC) and a buffer. For scaling, we sample the signals of Eqs. (10) and (11) and store them in the buffer as
where
where max() is the maximum of , min() is the minimum, and
To convert the scaled data to a unit circle, we extend Eqs. (12) and (13) to
Eqs. (18) and (19) have a respective phase difference of 90°, and differ in amplitude. We normalize the in-phase and quadrature signals by taking
Finally, we calculate the phase difference
III. EXPERIMENTAL SETUP AND RESULTS
We configured our experiment as shown in Figs. 1 and 3 above. We used a laser diode (LD) with a center wavelength of 1554.94 nm and sinusoidally modulated at approximately 15 MHz. We adjusted the quadrature-signal amplitudes by controlling the depth of the sinusoidal LD modulation. We coupled the LD output into a fiber-based optical interferometer composed of an optical coupler, a collimator, and a target mirror mounted on a motorized stage. The interfered signal was converted into an electrical signal by a photodetector (PD) and demodulated into quadrature signals through a mixer and a low-pass filter. These signals were collected by a data acquisition (DAQ) system composed of two 12-bit ADCs with 1-kHz sampling speed (M2i.3021, Spectrum Ins.). Finally, the proposed compensation method was implemented using LabVIEW software for
Figure 5 shows the experimental results using our proposed method, as compared to the previous method that only uses scaling. As shown in Fig. 5(a), amplitude fluctuation of the signal including dc offset has disappeared through the proposed scaling method. In Fig. 5(b), and with only scaling had a phase difference of about 45°, whereas and had a phase difference of almost 90° after fitting; the corresponding Lissajous pattern of and was much more circular in Fig. 5(c), since the phase difference was closer to 90°. Figure 5(d) shows the target displacement corrected using our proposed compensation, as compared to the previous method and to the ideal case. The displacement after both correction schemes still showed a small fluctuation, from about 5 nanometers using scaling only down to 2 nanometers using our proposed method of fitting plus scaling. We believe that the remaining fluctuation is most likely due to system noise, caused by high power modulation and external disturbances [2].
We proposed a simple quadrature-detection-error-compensation method and demonstrated this approach experimentally. We normalized the in-phase and quadrature signals so that the maximum and minimum amplitudes were mapped to +1 and −1, followed by fitting the scaled signal to a unit circle through signal processing. Our experimental results showed that we reduced the quadrature detection error from 5 nanometers down to 2 nanometers, with the residual error likely caused by system noise from high power modulation.