The exposure of humans to radiofrequency fields from hand-held and body-mounted wireless communication devices is assessed by measuring the specific absorption rate (SAR). The SAR is the initial time derivative of temperature in the human tissue of the head or body in kelvins per second as Eq. (1) [1].
The term
where
The SAR System validation test should be conducted before starting a SAR measurement procedure for verifying that the system operates within its specifications. The test setup for validation consists of a flat phantom and a system validation source. In the SAR system validation, the measured 1 g and/or 10 g averaged SAR value produces the numerical reference peak special-average SAR values in IEC 62209 [2, 3]. The uncertainty of the power delivered to the validation source is kept as low as possible. This requires the use of a test setup with directional couplers and power meters during the system check, as shown in Fig. 1 [2, 3].
The power source setup is also used for the probe calibration. The expanded uncertainty of the SAR measurement system is approximately 30%. The power drift in the validation test or in the waveguide for probe calibration is around 5%, which makes up a substantial part of the total uncertainty parameters. Fig. 2 shows the power drift without Automatic Input Power Level Control (AIPLC), which is the common phenomenon in the normal operation conditions with the usual signal generator and power amplifier.
The RF input power for system validation in Fig. 1 is controlled manually during the test process; thus, it requires more than 1 hour to reach the stable state to be used for a validation test. The IEC 62206 Part 1 and Part 2 require a power drift less than 5% in normal operating condition and the standard uncertainty by this power drift is 2.89%. The standard uncertainty (
II. SIGNAL SOURCE FOR THE VALIDATION SYSTEM AND AIPLC ALGORITHM
The wiring schematic diagram of the signal source for SAR validation system is shown in Fig. 3. A computer controls both the signal generator and the power meter automatically through a GPIB cable and its board. It also controls the amplifier through the RS232C cable.
The equipment is controlled by the programmed input power level control system for the stable input power level, whose algorithm is described in Fig. 4.
The AIPLC algorithm starts with clicking of the start button in Fig. 4, after setting up the equipment as shown in Fig. 3, and follows the procedure described hereafter.
III. RF OUTPUT POWER MEASUREMENTS WITH AIPLC
Figs. 5 and 6 show the output RF power variations with respect to operation time when the AIPLC is applied to the amplifier or to the signal generator. The power variation obtained by controlling the gain of the amplifier with AIPLC was about ±0.125 dB, and the power variation obtained by controlling the output of signal generator was less than ±0.05 dB.
AIPLC can be applied when calibrating the probe conversion factors. The conversion factor in tissue-equivalent liquid is determined by generating locally known field values inside the tissue with analytical fields in waveguides. The use of this method for calibrating probes in lossy liquids must take into account the net RF power dissipated in the waveguide by accurate measurements.
This requirement implies precise measurement of the incident power, reflected power, and reflection coefficient at the waveguide input port [6].
Here,
The variation range of the conversion factors obtained without AIPLC application was from –1.17% to 1.79% and the standard deviation was 0.13.
On the other hand, the variation range of the conversion factors with AIPLC was reduced from ±0.7% to ±0.28%, and the standard deviation was 0.04. The conversion factors with and without the application of AIPLC are given in Table 1.
[Table 1.] The probe CFs with and without the application of AIPLC
The probe CFs with and without the application of AIPLC
The required power drift of RF input power in validation and calibration process should be less than 5% according to IEC 62209-1. Currently, the input power for system validation or probe calibration is controlled manually during the process, so a long time is required to reach to the stable power needed for use in testing that will induce the least uncertainty. The AIPLC concept was suggested and implemented in the SAR measurement system. The application of AIPLC for validation of the SAR measurement system and for probe calibration process helps to maintain a stable RF output power from the start of the test by controlling the signal generator output or amplifier gain, which reduces the uncertainty due to RF power drift from 2.89% to 0.67%.