The rapid advances in communication technologies have led to the use of the wearable wireless body area network (WBAN) for various applications such as medical devices, police and military agencies, sports training, entertainment, and wearable computing etc. [1]. The antennas used for WBAN require a small size, a low human body effect, and a low specific absorption ratio (SAR) [2]. The human body has a high dielectric constant with a high loss tangent and low conductivity at the microwave frequency band. Therefore, the gain and radiation efficiency of an antenna can be reduced when an antenna is operated on or within the human body.
Recently, much work has been done to investigate the on-body communication channel at the Industrial Scientific Medical (ISM) band [3]~[5]. In a WBAN system, multipath propagation can occur due to reflections from the surrounding environment and the body parts. In addition, multipath fading can occur in response to the large relative movements of body parts, shadowing, polarization mismatch, and scattering by the body and the surrounding environment [3]. This harsh WBAN communication environment has been addressed by the recent proposal of various diversity techniques [6]. The placement of two antennas close to each other requires an isolation between the antennas that is sufficiently high to minimize the mutual effects.
This paper proposes a miniaturized high isolation diversity antenna for wearable WBAN applications in the ISM band (2.4~2.485 GHz). A planar inverted-F antenna (PIFA) is used to achieve a compact size of an antenna for on-body communication. Two antenna elements are placed on the same substrate with a small ground element in order to improve the performance and overcome multipath fading. The performance parameters of the proposed diversity antenna placed on the human body tissues, including the S-parameter characteristics, radiation patterns and mean effective gain (MEG), MEG ratio, envelope correlation coefficient (ECC), and specific absorption ratio (SAR), are analyzed through a simulation and measurements by using a semi-solid flat phantom that has electrical properties equivalent to those of a whole human body. The proposed diversity antenna was designed and analyzed using the Ansys HFSS v14 software [7]. The details of the antenna design and the experimental results are presented and discussed in the following sections.
Ⅱ. Antenna Design and Performance
The configuration of the proposed diversity antenna is shown in Fig. 1(a). The proposed diversity antenna consists of PIFAs, which are placed near the corners of the top edge of the ground plane. A T-shaped isolator is then placed between the two antenna elements. The two antenna elements of the diversity antenna are symmetrically placed with respect to the y-axis. Each antenna element has a dimension of 13 mm×10 mm×1.5 mm and is fed by a 50 Ω coaxial cable. Fig. 1(b) shows the structure of a T-shaped isolator that is used to enhance the isolation characteristic at the ISM band. The ground plane is composed of a 1-mm-thick FR-4 substrate with a relative dielectric constant of 4.4 and a size of 30 mm×30 mm.
Antenna performance when the antenna is placed on a human body is analyzed by running simulations using a human body flat phantom (200 mm×270 mm×60 mm) with
A parametric study of the proposed antenna is shown in Fig. 2. Figs. 2(a) and (b) show the simulated
The simulated
The fabricated antenna and the semi-solid flat phantom are shown in Figs. 5(a) and (b), respectively. The relative dielectric constant and conductivity of the fabricated semi-solid phantom [8], measured using an Agilent 8570E dielectric probe kit and an 8719ES network analyzer, are shown in Fig. 6. The phantom with a dimension of 200 mm×270 mm×60 mm is used to measure the
Fig. 8(a) and (b) show the measured 3D radiation patterns of the proposed antenna placed on the semisolid flat phantom. The measured peak gain of the antenna elements # 1 and # 2 is ?0.43 dBi and ?0.54 dBi at 2.45 GHz, respectively.
The performance of the proposed diversity antenna is evaluated by analyzing the key performance parameters such as the ECC, MEG, and MEG ratio. For diversity and multiple-input multiple-output (MIMO) applications, the correlation between the signals, received at the same side of a wireless link by the involved antenna, is an important figure of merit for the whole system. The ECC is normally used to evaluate the diversity capability of multi antenna system. This parameter should preferably be computed from 3-D radiation patterns [9]. A good diversity gain can be obtained when the ECC is less than 0.5. Fig. 9 shows the ECC characteristics computed from the measured radiation pattern. The ECC of the two antennas is maintained below 0.1 over the whole frequency band. This leads us to expect good performance in terms of diversity.
[Table 1.] ECC, MEG, and MEG ratio, and actual diversity gain of the proposed diversity antenna.
ECC, MEG, and MEG ratio, and actual diversity gain of the proposed diversity antenna.
The performance of the proposed diversity antenna, including the ECC, MEG, and MEG ratio, and actual diversity gain, is summarized in Table 1. The information in Table 1 shows that the received signals satisfy the conditions suggested in [9], such that:
When a uniform propagation environment is assumed, the MEG ratio is almost unity, indicating that the mean power delivered from the two ports are almost the same.
The SAR is an essential factor for consideration when the antenna is operated on or inside the human body. The SAR is measured using the ESSAY system [10] at the Radio Research Agency of Korea, as shown in Fig. 10. The proposed antenna is excited by a signal generator. Fig. 10 shows the measured SAR distributions of the proposed diversity antenna placed 5 mm outside of the liquid flat phantom, which has a dimension of 300 mm×200 mm×200 mm. It is filled with a liquid having electrical properties equivalent to those of the human body tissue (
The FCC of the United States requires that the SAR values should be below 1.6 W/kg over a volume of 1 gram of tissue. Delivery of an input power of 250 mW, which is the input power normally used for the SAR measurement of mobile application devices, results in the maximum SAR values obtained for the antenna elements #1 and #2, at 1.801 W/kg and 1.637 W/kg (1 g tissue), as shown in Figs. 11(a) and (b), respectively. Portable devices using 2.45 GHz applications, such as Bluetooth and Zigbee, operate below 100 mW (0.7204 W/kg for antenna element #1 and 0.6548 W/kg for antenna element #2 when the input power is 100 mW), so the maximum SAR values of the antenna are sufficiently low to satisfy the SAR limitation (1.6 W/kg for partial- body exposure) in practical use [8].
This paper proposes a miniaturized diversity antenna with high isolation for wearable WBAN applications that overcomes the multipath fading caused by large relative movements of the body parts. The proposed diversity antenna has a dimension of 30 mm×30 mm×2.5 mm. The isolation performance is improved by using a T-shaped isolator, which is placed between the two antenna elements. The performance of the proposed diversity antenna, including bandwidth, MEG ratio, ECC, and SAR distribution, is sufficient for use in WBAN applications. In addition, the antenna is well suited for onbody applications, owing to its compact size.